ChromaTwin

Gradient method transfer · Practical Pharma Science™

Basis · Measured dwell volume
M

Worked Reference Transfer

Agilent 1260 at 0.60 mL dwell to a Waters Alliance at 1.20 mL, 1.0 mL/min. Every figure below is produced by the same engine the tool runs — nothing here is typed in by hand.

Source t_D
0.600 min
0.60 mL ÷ 1.0 mL/min
Target t_D
1.200 min
1.20 mL ÷ 1.0 mL/min
Δt offset
+0.600 min
≡ +0.600 mL
Leading hold
2.00 min
≥ Δt, so shift is permitted

Source table beside the adjusted table

#Source time%BAdjusted time%BChange
10.005.00.005.0unchanged
22.005.01.405.0−0.60 min
320.0095.019.4095.0−0.60 min
425.0095.024.4095.0−0.60 min
525.105.024.505.0−0.60 min
630.005.030.005.0unchanged
What happened
The 2.00 min leading hold is longer than the 0.60 min offset, so the hold absorbs the cut: every programmed time comes forward by 0.60 min, and the final hold is extended by 0.60 min so the target still delivers the original re-equilibration volume. The run length from injection is unchanged at 30.00 min.

What it was worth

Run unchanged on the Waters, the gradient would have reached the column head 0.60 min late — 0.60 mL of mobile phase. Through the ramp that displaces the profile by 3.0 %B; across the wash-out step it is momentarily 90 %B, because the step itself arrives late. Every peak eluting under gradient control moves with it.

Compensated, the adjusted method reproduces the source column-head profile to within 0.00 %B, with a rounding residual of 0.000 min (0.000 mL). That is the whole claim, and it is checked on every build by npm run verify.