Gradient Method Transfer

Translate a gradient table between instruments with different measured dwell volumes. Closed-form arithmetic: it corrects the gradient delay, and nothing else.

Source V_D
0.60mL
t_D = 0.600 min
Target V_D
1.20mL
t_D = 1.200 min
Ξ”V dwell
+0.600mL
target minus source
Ξ”t delay offset
+0.600min
at 1.00 mL/min
Compensation
Initial hold shortened
6 steps rewritten

Configuration



Source gradient

#Time (min)%B
1
2
3
4
5
6

Adjusted gradient table β€” target

6 steps Β· 30.00 min from injection
Initial hold shortened
Every programmed time is brought forward by Ξ”t; the leading isocratic hold absorbs the cut. Composition at the cut point is 5.0 %B, which becomes the new t = 0 entry.
Residual after compensation
Programmed to 0.01 min resolution, the adjusted method reproduces the source composition profile at the column head to within 0.00 %B (timing residual 0.000 min ≑ 0.000 mL). Run unchanged, the target's gradient would have reached the column head 0.60 min late β€” displacing the profile by 3.0 %B through the gradient ramp.
#Time (min)%BSegment
10.005.0start
21.405.0hold
319.4095.0ramp ↑
424.4095.0hold
524.505.0step
630.005.0hold

Composition at the column head

Minutes from injection, against %B arriving at the column head. The compensated trace should sit on the source trace β€” that is the whole test.

Hardware compensation matrix

1Source gradient delay timet_D,src = V_D,src / F = 0.600 mL Γ· 1.000 mL/min0.600 min
2Target gradient delay timet_D,tgt = V_D,tgt / F = 1.200 mL Γ· 1.000 mL/min1.200 min
3Dwell volume differenceΞ”V = V_D,tgt βˆ’ V_D,src+0.600 mL
4Gradient delay offsetΞ”t = Ξ”V / F = t_D,tgt βˆ’ t_D,src+0.600 min
5Leading isocratic hold on sourceH = last programmed time at which %B is still 5.0 %2.000 min
6Feasibility testH β‰₯ Ξ”t β†’ 2.000 β‰₯ 0.600. The hold is long enough to absorb the cut.passed
7Compensation appliedEvery programmed time is brought forward by Ξ”t; the leading isocratic hold absorbs the cut.βˆ’0.60 min
8Rounding to instrument resolutionProgrammed times rounded to 0.01 min; the residual expressed as a volume at the method flow rate.0.000 min ≑ 0.000 mL
9VerificationMaximum difference between the source and adjusted column-head profiles, sampled at 1500 points.0.00 %B