Measuring the Dwell Volume

Every number ChromaTwin produces rests on two measured volumes. This is the ramp-extrapolation method, which is the one to use when the answer has to survive review.

  1. 1
    Take the column out of the flow path
    Replace it with a zero-dead-volume union. Leave the injector, tubing and detector exactly as the method runs them β€” you are measuring this system, not an idealised one.
  2. 2
    Prepare the tracer
    Line A: water. Line B: water containing 0.1 % (v/v) acetone β€” or 10 mg/L uracil. Degas both.
  3. 3
    Set the detector
    Flow at the method flow rate. Detection at 265 nm for acetone, 254 nm for uracil. 10 Hz or faster, no reference wavelength.
  4. 4
    Run the ramp
    Equilibrate at 0 %B, then programme a linear ramp 0 β†’ 100 %B over 10 min and hold 5 min. Do not inject.
  5. 5
    Read the intercept
    Extrapolate the flat baseline forward and the linear rising portion backward. Their intersection is tD.
  6. 6
    Convert to a volume
    Subtract the injector-to-detector tubing volume divided by flow if it exceeds ~1 % of tD. Then VD = tD Γ— F.
  7. 7
    Repeat, and re-measure when the plumbing changes
    Run in triplicate and report the mean. Re-measure after any change to mixer, tubing, needle seat or pump head β€” those are the parts the number is made of.
Cross-check
The step method gives the same answer: step 0 β†’ 100 %B and take the time at 50 % of the plateau height. Agreement within 0.05 min between the two methods is a good sign the measurement is sound.

Typical dwell volumes

For orientation only β€” to sanity-check a measurement, never to replace one. Two nominally identical systems routinely differ by several tenths of a millilitre, because mixer, needle seat and tubing all count toward the number.

SystemTypical V_D (mL)t_D at 1.0 mL/min
Agilent 1260 Infinity II β€” binary pump, standard mixer0.600.60 min
Agilent 1260 Infinity II β€” quaternary pump0.900.90 min
Agilent 1100/1200 β€” quaternary pump0.800.80 min
Agilent 1290 Infinity II β€” binary pump0.140.14 min
Waters Alliance 2695 / e26951.201.20 min
Waters ACQUITY UPLC H-Class0.400.40 min
Waters ACQUITY UPLC I-Class β€” binary0.120.12 min
Shimadzu Nexera X2 β€” LC-30AD binary0.550.55 min
Shimadzu Prominence LC-20AT β€” quaternary0.900.90 min
Thermo Vanquish Horizon β€” binary0.350.35 min
Thermo UltiMate 3000 β€” quaternary0.650.65 min