Measuring the Dwell Volume
Every number ChromaTwin produces rests on two measured volumes. This is the ramp-extrapolation method, which is the one to use when the answer has to survive review.
- 1Take the column out of the flow pathReplace it with a zero-dead-volume union. Leave the injector, tubing and detector exactly as the method runs them β you are measuring this system, not an idealised one.
- 2Prepare the tracerLine A: water. Line B: water containing 0.1 % (v/v) acetone β or 10 mg/L uracil. Degas both.
- 3Set the detectorFlow at the method flow rate. Detection at 265 nm for acetone, 254 nm for uracil. 10 Hz or faster, no reference wavelength.
- 4Run the rampEquilibrate at 0 %B, then programme a linear ramp 0 β 100 %B over 10 min and hold 5 min. Do not inject.
- 5Read the interceptExtrapolate the flat baseline forward and the linear rising portion backward. Their intersection is tD.
- 6Convert to a volumeSubtract the injector-to-detector tubing volume divided by flow if it exceeds ~1 % of tD. Then VD = tD Γ F.
- 7Repeat, and re-measure when the plumbing changesRun in triplicate and report the mean. Re-measure after any change to mixer, tubing, needle seat or pump head β those are the parts the number is made of.
Cross-check
The step method gives the same answer: step 0 β 100 %B and take the time at 50 % of the plateau height. Agreement within 0.05 min between the two methods is a good sign the measurement is sound.Typical dwell volumes
For orientation only β to sanity-check a measurement, never to replace one. Two nominally identical systems routinely differ by several tenths of a millilitre, because mixer, needle seat and tubing all count toward the number.
| System | Typical V_D (mL) | t_D at 1.0 mL/min |
|---|---|---|
| Agilent 1260 Infinity II β binary pump, standard mixer | 0.60 | 0.60 min |
| Agilent 1260 Infinity II β quaternary pump | 0.90 | 0.90 min |
| Agilent 1100/1200 β quaternary pump | 0.80 | 0.80 min |
| Agilent 1290 Infinity II β binary pump | 0.14 | 0.14 min |
| Waters Alliance 2695 / e2695 | 1.20 | 1.20 min |
| Waters ACQUITY UPLC H-Class | 0.40 | 0.40 min |
| Waters ACQUITY UPLC I-Class β binary | 0.12 | 0.12 min |
| Shimadzu Nexera X2 β LC-30AD binary | 0.55 | 0.55 min |
| Shimadzu Prominence LC-20AT β quaternary | 0.90 | 0.90 min |
| Thermo Vanquish Horizon β binary | 0.35 | 0.35 min |
| Thermo UltiMate 3000 β quaternary | 0.65 | 0.65 min |